Kilovoltik is semi-professional muscle stimulator which is using the nerves as part of an electric circuit.
The device operates at 24V with user controllable current in range <0-45> mA. The user can select either AC or DC; both modes have multiple signal profiles and adjustable frequency.
The device has built-in battery and can't operate while charging - there is only one (overloaded) connector both for electrodes (to be placed on the subject) and the power cable.
Prerun check:
Disconnect power cable
Connect electrodes to the subject:
One electrode under an joint (pod kloubem)
Second over the same joint (nad kloubem)
Connect the cable to the machine
Power ON:
Setup:
Select wave profile (most left group)
Check the “Error” diode NOT to shine
For newbies, the TENS(n) is good 1st choise
Select the frequency (center; if applicable)
How often the pulse should go (Hz)
Select wave length (right, if applicable)
How long the wave should be (obviously this mutiplied by grequency can't be > 1s)
If the Error LED is shining:
Check if the subject is corectly connected
Especially girl's skin have higher resistance due to usage of makeups, you can clean them using pure ethanol (flask on the shelf with “Etoh” label)
Switch electrodes (or use “crocodiles” for better attachment to the skin)
Operation:
Use “STOP” button under the current controls to quickly nullify the current.
Do NOT raise the current over 35 mA for long time, it can harm the nerves. Do NOT go over 42 mA even for short time on small muscles, over short time on big ones.
Power OFF:
Its OK, you can breathe